**** UBC ECE E-HUB **** **** RIGOL M302DAQ LOG BOOK **** **** Please submit any new entries by e-mail to: robertor [at] ece.ubc.ca **** Entries include tips, bugs, damage reports, custom improvements, etc. DOCUMENT HISTORY Created: 26/May/2025 Last Change: 26/May/2025 Version: 1.00.01 version digits indicate # updates of type: [ description, maintenance, user ] **** STATUS: ------------------------------------------------------------------------------------ 1. INSTRUMENT DESCRIPTION **** 1.1 ORIGIN Bought on March 2024 from ECE funds for M1022 **** 1.2 CAPABILITIES The M302 is a Data Acquisition Mainframe with 5 slots that can host DMM, MUX, digital IO, and matrix modules. With the right modules it is possible to reach 320 channels. The DMM measures voltage, current, resistance, frequency, test diodes, temperature, and can scan up to 256 channels. MC3065 6.5 DMM module: | Measurement Type | Ranges | Resolution | | ----------------------| --------------------------------- | ------------------------- | | DC Voltage | 100 mV to 300 V | Up to 100 nV | | AC Voltage (True RMS) | 100 mV to 300 V | Up to 1 uV | | DC Current | 10 uA to 3 A | Up to 10 nA | | AC Current (True RMS) | 10 uA to 3 A | Up to 100 nA | | 2-wire Resistance | 100 ohm to 100 Mohm | Up to 100 uohm | | 4-wire Resistance | 1 ohm to 10 Mohm | High precision, low drift | | Frequency/Period | Up to 300 kHz | | | Diode Test | Yes | | | Temperature | Via thermocouples (J, K, T, etc.) | oC, oF, K | MC3120 MUX module: | Parameter | Value | | -------------------------------------------------| --------------------------------------------- | | Number of Channels | 20 2-wire channels | | Switching Type | Multiplexer (1 of 20 to common) | | Maximum Voltage (Ch-to-Channel or Ch-to-Chassis) | +/-300 V DC or RMS AC.. | | Maximum Current | 2 A continuous per channel | | Maximum Power per Channel | 60 W or 62.5 VA | | Relay Type | Electromechanical relays | | Switching Speed | ~50 ch/s (depends on scanning configuration) | | Resistance (Closed Channel) | < 1 ohm typical | | Isolation Between Channels | > 10^9 ohm typical (1 Gohm) | M3TB20 MUX Terminal block for the MC3120 MUX: Interface between the signal wires and the MUX **** 1.3 CONFIGURATION WE HAVE M302 (MAINFRAME) + MC3065 (DMM) + MC3120 (20CH-MUX) + M3TB20 (MUX TERMINAL) M302 SN: MM3A261000077 ECE: 102114 Instrument URL: MAC ADDRESS: MC3065 SN: MM3C60300004 ECE: 102115 MC3120 SN: MM3L255100252 ECE: 10216 M3TB20 SN: TM3G252900109 ECE: 102117 Options: **** 1.4 MANUALS https://people.ece.ubc.ca/~eng-services/files/eHUB/Bench3-Parametric/ ------------------------------------------------------------------------------------ 2. INSTRUMENT MAINTENANCE ------------------------------------------------------------------------------------ 3. USER ENTRIES 26-May-2025 Roberto: Downloaded "UltraAcquire data acquisition PC software for power users" to the Manuals folder. Did not install nor try.