This calibration procedure applies to the 34465A and 34470A only.
Review Test Considerations and Flatness Calibration Considerations before performing any calibration.
Configuration: AC Current
All AC calibration procedures use the 3 Hz bandwidth measurement filter.
![]() |
Each current and frequency takes approximately 2.5 seconds to complete. |
Input | Instrument | |
---|---|---|
Current | Frequency | Range |
1.5 A | 5 kHz, 7.5 kHz, 9.7 kHz, 10 kHz, 220 Hz, 1 kHz | 3 A |
1 A | 5 kHz, 7.5 kHz, 9.7 kHz, 10 kHz, 220 Hz, 1 kHz | 1 A |
100 mA | 5 kHz, 7.5 kHz, 9.7 kHz, 10 kHz, 220 Hz, 1 kHz | 100 mA |
10 mA | 5 kHz, 7.5 kHz, 9.7 kHz, 10 kHz, 220 Hz, 1 kHz | 10 mA |
1 mA | 5 kHz, 7.5 kHz, 9.7 kHz, 10 kHz, 220 Hz, 1 kHz | 1 mA |
100 µA | 5 kHz, 7.5 kHz, 9.7 kHz, 10 kHz, 220 Hz, 1 kHz | 100 µA |