The ADC and zero calibration adjustment procedure must have been recently performed before beginning gain calibration. Gain calibrations can be made using either the front or rear input (not available on the 34460A) terminals. On the 34461A/65A/70A, make sure the Front/Rear switch matches the terminals in use.
![]() |
Never turn off the instrument during a gain or flatness calibration. This may cause calibration memory for the present function to be lost. |
Gain calibration can be accomplished using the following input values.
Function | Range | Valid Amplitude Input Values |
---|---|---|
DC Volts | 100 mV to 100 V | 0.9 to 1.1 x Full Scale |
1000 V | 450 to 1000 V | |
DC Current | 100 µA to 1 A | 0.9 to 1.1 x Full Scale |
3 A | 1.5 A to 2.5 A | |
10 A | 1.8 A to 10 A | |
Ohms (2W and 4W) | 100 Ω to 100 MΩ | 0.9 to 1.1 x Full Scale |
Frequency | Auto | Input > 100 mV rms, 9.9 kHz to 110 kHz |
AC Current (rms) [1] | 100 µA to 1 A | 0.9 to 1.1 x Full Scale |
3 A | 0.9 A to 2 A | |
10 A | 2 A to 10 A | |
AC Volts (rms) [1] | 100 mV to 10 V | 0.9 to 1.1 x Full Scale |
100 V | 40 V to 110 V | |
750 V | 400 to 750 V |
[1] Valid frequencies are nominal frequencies ±1%.