If TRIGger:DELay:AUTO is on, the instrument selects the trigger delay for you, as shown in the tables below.
![]() |
All frequency and period measurements have an automatic trigger delay of 1 second. The continuity and diode functions ignore the trigger delay. The 2-wire and 4-wire temperature measurements use the corresponding 2-wire and 4-wire resistance measurement delays. |
PLC |
0.0011 |
0.0021 |
0.0061 |
0.02 |
0.062 |
0.2 |
1 |
10 |
100 |
---|---|---|---|---|---|---|---|---|---|
Aperture (60 Hz Power) |
20 µs1 |
40 µs1 |
100 µs1 |
0.3 ms |
1 ms2 |
3 ms |
16.7 ms |
0.167 s |
1.67 s |
Aperture (50 Hz Power) |
20 ms |
0.2 s |
2 s |
||||||
Range |
Delay |
||||||||
All |
100 µs |
130 µs |
160 µs |
1 34465A/70A with DIG option only.
234465A/70A only.
PLC |
0.0011 |
0.0021 |
0.0061 |
0.02 |
0.062 |
0.2 |
1 |
10 |
100 |
---|---|---|---|---|---|---|---|---|---|
Aperture (60 Hz Power) |
20 µs1 |
40 µs1 |
100 µs1 |
0.3 ms |
1 ms2 |
3 ms |
16.7 ms |
0.167 s |
1.67 s |
Aperture (50 Hz Power) |
20 ms |
0.2 s |
2 s |
||||||
Range |
Delay |
||||||||
All |
1 ms |
1.5 ms |
1 34465A/70A with DIG option only.
2 34465A/70A only.
PLC |
0.0011 |
0.0021 |
0.0061 |
0.02 |
0.062 |
0.2 |
1 |
10 |
100 |
---|---|---|---|---|---|---|---|---|---|
Aperture (60 Hz Power) |
20 µs1 |
40 µs1 |
100 µs1 |
0.3 ms |
1 ms2 |
3 ms |
16.7 ms |
0.167 s |
1.67 s |
Aperture (50 Hz Power) |
20 ms |
0.2 s |
2 s |
||||||
Range |
Delay |
||||||||
100 Ω |
80 µs |
100 µs |
130 µs |
||||||
1 kΩ |
110 µs |
130 µs |
160 µs |
||||||
10 kΩ |
130 µs |
160 µs |
190 µs |
||||||
100 kΩ |
540 µs |
670 µs |
800 µs |
||||||
1 MΩ |
5 ms |
6 ms |
7.5 ms |
||||||
10 MΩ |
60 ms |
70 ms |
84 ms |
||||||
100 MΩ |
|||||||||
1 GΩ2 |
1 34465A/70A with DIG option only.
2 34465A/70A only.
PLC |
0.0011 |
0.0021 |
0.0061 |
0.02 |
0.062 |
0.2 |
1 |
10 |
100 |
---|---|---|---|---|---|---|---|---|---|
Aperture (60 Hz Power) |
20 µs1 |
40 µs1 |
100 µs1 |
0.3 ms |
1 ms2 |
3 ms |
16.7 ms |
0.167 s |
1.67 s |
Aperture (50 Hz Power) |
20 ms |
0.2 s |
2 s |
||||||
Range |
Delay |
||||||||
100 Ω |
1 ms |
1.5 ms |
|||||||
1 kΩ |
|||||||||
10 kΩ |
|||||||||
100 kΩ |
|||||||||
1 MΩ |
10 ms |
15 ms |
|||||||
10 MΩ |
100 ms |
||||||||
100 MΩ |
|||||||||
1 GΩ2 |
1 34465A/70A with DIG option only.
2 34465A/70A only.
|
Filter (Hz) |
||
---|---|---|---|
Range |
3 |
20 |
200 |
All |
2.5 s |
0.625 s |
0.025 s |
|
Filter (Hz) |
||
---|---|---|---|
Range |
3 |
20 |
200 |
All |
1.66 s |
0.25 s |
0.025 s |
The default delay for frequency and period is 1 s, regardless of aperture (gate time) setting.